News | Endoscopes | January 05, 2016

FDA-approved instructions include more rigorous protocol for pre-cleaning, manual cleaning and high-level disinfection procedures

January 5, 2016 — The U.S. Food and Drug Administration (FDA) posted a new Safety Communication titled, "Fujifilm Medical Systems U.S.A. Inc. Validates Revised Reprocessing Instructions for Model ED-530XT Duodenoscopes." 

As noted in FDA’s February 2015 Safety Communication, the complex design of duodenoscopes may impede effective reprocessing. Reprocessing is a detailed, multistep process to clean and disinfect or sterilize reusable devices. The FDA has been working with duodenoscope manufacturers as they modify and validate their reprocessing instructions to further enhance the safety margin of their devices and show with a high degree of assurance that their reprocessing instructions, when followed correctly, effectively clean and disinfect the duodenoscopes.

In May 2015, Fuji initiated testing to validate the revised reprocessing instructions and provided the cleaning validation reports to FDA in July 2015. Between July and October 2015, the agency continued to work with Fuji to clarify and ultimately confirm that their cleaning and high-level disinfection instructions met the agency’s expectations. In October 2015, the FDA notified Fuji that the validation data for the ED-530XT duodenoscope were acceptable and worked with the company as they developed revised instructions for use.

The revised instructions include a more rigorous protocol for pre-cleaning, manual cleaning and high-level disinfection procedures. They also incorporate the use of an additional disposable brush (Model WB1318DE) to manually clean the distal end of the scope. The agency reviewed the validation data and believes that when followed, the revised, validated reprocessing instructions demonstrate consistent and reliable cleaning and high-level disinfection of the Fuji ED-530XT duodenoscope.

For more information: www.fujimed.com


Additional Content

News | Radiographic Fluoroscopy (RF)

April 8, 2026 — Alpha Imaging, a Radon Medical Imaging company, will display its latest advancement in image-guided ...

Time April 08, 2026
arrow
News | ACR

February 6, 2024 — An update to the CT Colonography Reporting and Data System (C-RADS) has been published Jan. 30 in the ...

Time February 06, 2024
arrow
News | Endoscopes

December 23, 2019 — The U.S. Food and Drug Administration cleared for marketing in the U.S. the first fully disposable ...

Time December 23, 2019
arrow
News | Endoscopes

December 23, 2019 — Boston Scientific Corporation announced U.S. Food and Drug Administration (FDA) 510(k) clearance of ...

Time December 23, 2019
arrow
News | Endoscopes

October 22, 2019 — OmniVision Technologies Inc. announced that its OV6948 camera module is the winner of the Guinness ...

Time October 22, 2019
arrow
Subscribe Now