News | Quality Assurance (QA) | March 08, 2016

Alternative standard based on ACR Digital Mammography Quality Control Manual

FDA, alternative standard, full field digital mammography, FFDM, qualtity assurance QA

March 8, 2016 — The U.S. Food and Drug Administration (FDA) announced the approval of an alternative digital mammography quality assurance (QA) program that mammography facilities may use instead of the program recommended by the image receptor manufacturer. The alternative standard is based on the American College of Radiology (ACR) Digital Mammography Quality Control Manual.

The original standard is 21 CFR 900.12(e)(5)(vi), which states:

900.12(e)(5)(vi): Quality control tests--other modalities. For systems with image receptor modalities other than screen-film, the quality assurance program shall be substantially the same as the quality assurance program recommended by the image receptor manufacturer, except that the maximum allowable dose shall not exceed the maximum allowable dose for screen-film systems in paragraph (e)(5)(vi) of this section.

The approved alternative is:

900.12(e)(5)(vi): Quality control tests--other modalities. For full-field digital mammography systems without advanced imaging capabilities, the quality assurance program shall be substantially the same as the quality assurance program recommended by the ACR Digital Mammography Quality Control Manual when used with the ACR Digital Mammography Phantom, except that the maximum allowable dose shall not exceed the maximum allowable dose for screen-film systems in paragraph (e)(5)(vi) of this section.

Any facility may avail itself of the approved alternative standard for the described imaging systems.

For more information: www.fda.gov


Additional Content

News | Women's Health

April 16, 2026 – GE HealthCare has expanded its collaboration with DeepHealth, Inc., a wholly-owned subsidiary of RadNet ...

Time April 20, 2026
arrow
News | Breast Imaging

April 15, 2026 — QT Imaging Holdings, Inc. has launched its QTI Imaging-Olea Viewer, developed in collaboration with ...

Time April 15, 2026
arrow
News | Breast Imaging

April 1, 2026 — QT Imaging Holdings has released its latest image reconstruction software update, version 4.5.0. This ...

Time April 02, 2026
arrow
News | Ultrasound Imaging

March 30, 2026 — Butterfly Network, Inc. has received clearance from the U.S. Food and Drug Administration (FDA) for a ...

Time April 01, 2026
arrow
News | Breast Imaging

March 30, 2026 — Each year, the Alumni Association at the University of Missouri-Kansas City, recognizes the ...

Time March 31, 2026
arrow
Subscribe Now