News | Colonoscopy Systems | February 08, 2018

Pentax Medical issued an Urgent Medical Device Correction and Removal notification on February 7, 2018. 

FDA Approves Updated Design for Recalled Pentax Duodenoscope

February 8, 2018 – The U.S. Food and Drug Administration (FDA) announced on February 7, 2018, that Pentax Medical’s updated design and labeling for its ED-3490TK Duodenoscope after the company issued an Urgent Medical Device Correction and Removal notification that same day.

The voluntary recall was put in place after Pentax informed customers in January 2017 about a potential issue associated with the distal cap of the ED-3490TK and began a program to inspect all customers' ED-3490TK duodenoscopes.

The recall is to replace the forceps elevator mechanism, O-ring seal and distal end cap, and to update the operation manual to recommend annual maintenance. The design changes are intended to reduce the potential for leakage of patient fluids into the closed elevator channel and under the distal cap.

As noted in the FDA's February 2015 Safety Communication, the complex design of duodenoscopes may impede effective reprocessing. Reprocessing is a detailed, multistep process to clean and disinfect or sterilize reusable devices. The FDA has been working with duodenoscope manufacturers as they modify and validate their reprocessing instructions to further enhance the safety margin of their devices and show with a high degree of assurance that their reprocessing instructions, when followed correctly, effectively clean and disinfect the duodenoscopes.

In February 2016, Pentax issued updated, validated manual reprocessing instructions for the ED-3490TK Duodenoscope to replace those provided in the original device labeling. The FDA reviewed these updated reprocessing instructions and the validation data and continues to recommend that facilities using Pentax ED-3490TK duodenoscopes train staff on the updated instructions and implement them, if they have not already done so.

The reprocessing instructions for the recently cleared ED-3490TK have not changed since the February 2016 update.

For more information: Updated Status of Pentax Medical Duodenoscope Model ED-3490TK: FDA Safety Communication


Additional Content

News | ACR

February 6, 2024 — An update to the CT Colonography Reporting and Data System (C-RADS) has been published Jan. 30 in the ...

Time February 06, 2024
arrow
News | Colonoscopy Systems

A team of Johns Hopkins data researchers is studying the economic and safety implications associated with the devices used to perform colonoscopies as a disposable version inches closer to widespread availability.

Time August 26, 2019
arrow
News | Colonoscopy Systems

A new study finds that trends in colonoscopy rates did not fully align with the increase in colorectal cancer (CRC) in younger adults, adding to evidence that the rise in early onset CRC is not solely a result of more detection. The study is published early online in the Journal of Medical Screening.

Time August 06, 2019
arrow
News | Colonoscopy Systems

Check-Cap Ltd. has initiated its U.S. pilot study of the C-Scan system for prevention of colorectal cancer through detection of precancerous polyps, following Institutional Review Board (IRB) approval and full Investigational Device Exemption (IDE) application approval by the U.S. Food and Drug Administration (FDA). The first patients have ingested C-Scan, a preparation-free capsule, at the New York University School of Medicine.

Time April 15, 2019
arrow
News | Colonoscopy Systems

https://www.congress.gov/bill/115th-congress/house-bill/1298?q=%7B%22se… cancer care advocacy groups and medical societies are calling on Congress to pass the CT Colonography Screening for Colorectal Cancer Act (S.3465/HR 1298). The bipartisan bill introduced in the U.S. Senate September 19 would provide Medicare coverage for screening computed tomography colonography (CTC), commonly known as virtual colonoscopy.

Time September 27, 2018
arrow
Subscribe Now